Surface resistance tester ACL-380

Release:2020-06-03  Views:

#Acl-380 surface resistance tester


Brand: ACL Staticide origin: USA


Parameters:


Power supply: 9 V PP3 alkaline battery


Measuring voltage: 9 V


Temperature range: operating temperature 5 ℃ to 49 ℃


Storage temperature: - 15 ℃ to 60 ℃


Relative humidity: 0% to 90% (non dewing state)


Measurement: the order of 10


Transition point: 1 / 2 order logarithm (3.16 × 10N)


Conversion point accuracy: ± 1 / 2 level


Accuracy: ± 10%


Repeatability: ± 5%


Weight: 170 g


Dimensions: 130 × 70 × 25 (mm units)


Calibration procedure of acl-380 surface impedance meter


Acl-380 adopts 1% ultra thin film resistor and high-speed amplifier circuit, and its calibration procedure is based on the linear circuit technology of the product.


The conversion point of impedance value is 1 / 2 logarithm, i.e. 3.16 × 10N. The linear value of conversion point is ± 10% of the mean value per unit of each measurement.


Measuring instrument: decade resistance box, test electrode


The resistance of the resistance box needs to be 1 × 103 ohm to 999 × 106 megohm or 109. When measuring the resistance value greater than 109, CAD generated technologies are used, because it is difficult to verify the resistance with impedance value greater than 999 × 106 by using 9 V voltage drive.


Connect the two electrodes of the resistance box to the two parallel electrodes of acl-380, adjust the resistance to 1K position, the 103 LED of the meter should be on, and the calibration of other gears should follow the same principle. When measuring the impedance conversion point, press the switch and continuously adjust the impedance value of the impedance bridge, and observe the impedance value when the next LED lamp is normally on (the value at this time is the conversion impedance value).


#It can be equipped with test weight, which is an optional accessory.